Edit:    Click: 1057  Date: 2011-07-22

Pressure gauge pressure gauge accuracy is accuracy grade level or level of accuracy referred to, (GB/T1226-2001 "general gauge" instrument of national standards known as the precision level, JJG52-1999 "spring tube general gauge, Pressure vacuum gauge and vacuum gauge, "called meter accuracy level. manometer accuracy classes: 1.0; 1.6; 2.5; 4.0, the level of the nominal diameter of the instrument case should meet the following requirements.

Nominal shell diameter (mm) Accuracy class
40,50,60 40,50,60 2.5, 2.5,
100 100 1.6,2.5 1.6,2.5
100,150,160,200 100,150,160,200 0.4, 1.0,1.6 0.4, 1.0,1.6

 

Gauge level of accuracy and tolerances and their relationships following table two

Accuracy class Tolerance% (calculated according to the percentage range)
Zero (90~100)% Measurement limo 100)% Rest
With only pin Non-stop sales
1.0 1 Plus or minus 1 Plus or minus 1.6 Plus or minus 1
1.6 (1.5) 1.6 Plus or minus 1.6 Plus or minus 2.5 Plus or minus 1.6
2.5 2.5 Plus or minus 2.5 Plus or minus 4 Plus or minus 2.5
4.0   Plus or minus 4 Plus or minus 4 Plus or minus 4
Note: Use the 1.5 gauge tolerance by 1.6 calculations, the level of accuracy may change

 

In the reference operating conditions, pressure gauge accuracy level testing should include the following four items:
1 show error in the measurement range, showing the error should be less than allowed under Table II error.
2 return the error in the measuring range, return error shall not exceed the allowable provided in Table 2 the absolute value of the error.
3. Tap tap displacement case, the pointer shows changes in the value should be less than the amount specified in Table 2 allowed absolute error of 1 / 2.
4 pointer deflection stationary in the measuring range, the pointer deflection should be smooth, non-beating and stuck phenomenon